Biocompatibility Assessment of Synthetic Sling Materials for Female Stress Urinary Incontinence

  • Alex Gomelsky

      Affiliations

    • Department of Urology, Louisiana State University Health Sciences Center-Shreveport, Shreveport, Louisiana
    • Financial interest and/or other relationship with Novartis.
  • ,
  • Roger R. Dmochowski

      Affiliations

    • Department of Urologic Surgery, Vanderbilt University Medical Center, Nashville, Tennessee
    • Corresponding Author InformationCorrespondence: Department of Urologic Surgery, Vanderbilt University Medical Center, A-1302, Medical Center North, Nashville, Tennessee 37232 (telephone: 615-343-5602; FAX: 615-322-8990).
    • Financial interest and/or other relationship with Allergan, Novartis and Watson Pharmaceuticals.

Received 4 October 2006 published online 15 August 2007.

Purpose

We evaluated the performance and complications of currently available synthetic sling materials with a focus on in vitro and in vivo biocompatibility, and acceptance in the human body.

Materials and Methods

We reviewed the MEDLINE® database for relevant literature pertaining to various synthetic sling materials. The Food and Drug Administration regulations regarding the regulation and biocompatibility testing of synthetic meshes were also reviewed.

Results

Many synthetic meshes used for sling construction were introduced before rigorous Food and Drug Administration regulations were passed and, thus, some became associated with unique complications. Most meshes used in pubovaginal and mid urethral sling surgery are associated with high short-term success rates and relatively few intraoperative complications. Despite modifications and additives, slings constructed from polytetrafluoroethylene and polyethylene are poorly accepted by the human body. Flexible, macroporous, polypropylene meshes appear to integrate more completely with human tissue than other synthetic materials. However, multifilament and nonknitted polypropylene slings may integrate poorly.

Conclusions

The composition, weave and pore size of each material are unique. These properties are responsible for the strength and durability of the material, as well as the ultimate acceptance and incorporation in the human body. Each material should be individually evaluated and patients should be counseled appropriately before implantation.

Key Words: suburethral slings, urinary incontinence, prostheses and implants

Abbreviations and Acronyms: FDA, Food and Drug Administration, ISO, International Standards Organization, IVS, intravaginal sling, PE, polyethylene, PET, polyethylene terephthalate, PP, polypropylene, PTFE, polytetrafluoroethylene, TVT, tension-free vaginal tape

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 30.00 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0022-5347(07)01394-8

doi:10.1016/j.juro.2007.05.123

The Journal of Urology
Volume 178, Issue 4 , Pages 1171-1181, October 2007